Atrioventricular Synchrony Delivered by a Dual-Chamber Leadless Pacemaker System

James E. Ip, Mayer Rashtian, Derek V. Exner, Vivek Y. Reddy, Rahul N. Doshi, Nima Badie, Jordan R. Nevo, Aditya Goil, Pascal Defaye, Robert C. Canby, Maria Grazia Bongiorni, Morio Shoda, Gerhard Hindricks, Reinoud E. Knops

2024Circulation27 citationsDOIOpen Access PDF

Abstract

BACKGROUND: A dual-chamber leadless pacemaker system has been designed for atrioventricular synchronous pacing using wireless, beat-to-beat, implant-to-implant (i2i) communication between distinct atrial and ventricular leadless pacemakers. The atrioventricular synchrony achieved across various ambulatory scenarios has yet to be systematically evaluated. METHODS: A prospective, single-arm, unblinded, multicenter, international clinical trial of the leadless pacemaker system was conducted in patients with a conventional dual-chamber pacing indication enrolled from February 2022 to March 2023. Leadless pacemaker systems were implanted, and 12-lead Holter electrocardiographic recordings were collected 3 months after implantation over various postures/activities: sitting, supine, left lateral recumbent, right lateral recumbent, standing, normal walk, and fast walk. An independent Holter core laboratory performed a manual adjudication of the percent of atrioventricular synchronous beats using the standard 300-millisecond PR interval limit. Atrium-to-ventricle and ventricle-to-atrium i2i communication success rates were also assessed. Post hoc summary statistics describing the relationships between atrioventricular synchrony and i2i success, posture/activity, implantation indication, atrioventricular event, and heart rate were calculated. RESULTS: <0.001), exceeding both i2i values in 95% of patients. Atrioventricular synchrony was achieved in >95% of evaluable beats across all postures/activities, implantation indications, atrioventricular paced/sensed event combinations, and heart rate ranges (including >100 bpm). CONCLUSIONS: This dual-chamber leadless pacemaker system demonstrated atrioventricular synchrony in 98% of evaluable beats at 3 months after implantation. Atrioventricular synchrony was maintained across postures/activities and remained robust for heart rates >100 bpm.

Topics & Concepts

MedicineSingle chamberCardiologyDual (grammatical number)Atrioventricular nodeInternal medicineTachycardiaLiteratureArtCardiac pacing and defibrillation studiesCardiac Arrhythmias and TreatmentsECG Monitoring and Analysis
Atrioventricular Synchrony Delivered by a Dual-Chamber Leadless Pacemaker System | Litcius