Litcius/Paper detail

Combination of two complementary automated rapid assays for diagnosis of heparin‐induced thrombocytopenia (HIT)

Theodore E. Warkentin, Jo‐Ann I. Sheppard, James W. Smith, Na Li, Jane C. Moore, Donald M. Arnold, Ishac Nazy

2020Journal of Thrombosis and Haemostasis35 citationsDOIOpen Access PDF

Abstract

BACKGROUND: HIT diagnosis typically uses complementary diagnostic assays (eg, a PF4-dependent enzyme-immunoassay [EIA] and a platelet activation assay such as the serotonin-release assay [SRA]). OBJECTIVES: To determine whether the combination of two automated assays-a latex immunoturbidimetric assay (LIA) that evaluates competitive inhibition of a HIT-like monoclonal antibody and a chemiluminescence immunoassay (CLIA) for detecting anti-PF4/heparin IgG-optimizes diagnostic sensitivity while also yielding good specificity, particularly at high assay reactivities. PATIENTS/METHODS: We determined operating characteristics using combined LIA/CLIA results from a HIT observational trial (n = 430; derivation cohort) and 147 consecutive patients with HIT (n = 147; supplementary derivation cohort). We also evaluated 678 consecutive samples referred for HIT testing (replication cohort). LIA/CLIA reactivities were scored individually as "negative" (<1.00 U/mL, 0 points), "weak" (1.00-4.99 U/mL, 1 point), "moderate" (5.00-15.99 U/mL, 2 points) and "strong" (≥16.00 U/mL, 3 points), thus contributing up to 6 points (maximum) when LIA/CLIA results were combined. We also examined whether higher LIA/CLIA scores predicted presence of platelet-activating antibodies by conventional and modified (PF4- or PF4/heparin-enhanced) SRA. RESULTS: Combined LIA/CLIA testing yielded high diagnostic sensitivity (~99%) similar to EIA. Interpretation of LIA/CLIA results using the 6-point scale indicated progressively greater likelihood for the presence of platelet-activating antibodies with increasing scores (semi-quantitative reactivity). A LIA/CLIA score ≥ 4 points predicted the presence of platelet-activating antibodies by SRA or PF4-enhanced SRA with high probability (~98%). CONCLUSION: Combined LIA/CLIA testing optimizes diagnostic sensitivity, with progressively greater probability of detecting platelet-activating antibodies with higher assay reactivity that reaches 98% when both automated assays yield moderate or strong results.

Topics & Concepts

ImmunoassayHeparin-induced thrombocytopeniaMedicineChemiluminescent immunoassayPlateletAntibodyInternal medicineCohortMonoclonal antibodyPlatelet activationHeparinImmunologyHeparin-Induced Thrombocytopenia and ThrombosisPlatelet Disorders and TreatmentsIntramuscular injections and effects