Litcius/Paper detail

Analytical Transmission Electron Microscopy

Wilfried Sigle

2020Max Planck Institute for Plasma Physics23 citationsDOIOpen Access PDF

Abstract

▪ Abstract Chemical analysis at high spatial resolution is the domain of analytical transmission electron microscopy. Owing to rapid instrumental developments during the past decade, electron energy-loss spectroscopy offers now a spatial resolution close to 0.1 nm and an energy resolution close to 0.1 eV. This development has been accompanied by the introduction of numerous new techniques and methods for data acquisition and analysis, which are outlined in the present article. Recent results for a wide range of material systems are addressed. These comprise first-principles calculations, which have contributed to enormous progress in the calculation of near-edge fine structures, and fingerprinting methods, which are still important for the interpretation of experimental data.

Topics & Concepts

Electron energy loss spectroscopyResolution (logic)Range (aeronautics)Energy filtered transmission electron microscopyTransmission electron microscopyNanotechnologyMaterials scienceScanning transmission electron microscopyComputer scienceArtificial intelligenceComposite materialElectron and X-Ray Spectroscopy TechniquesAdvanced Electron Microscopy Techniques and ApplicationsIon-surface interactions and analysis
Analytical Transmission Electron Microscopy | Litcius