Suboptimal Inspiratory Flow Rates With Passive Dry Powder Inhalers: Big Issue or Overstated Problem?

Jeffry G. Weers

2022Frontiers in Drug Delivery12 citationsDOIOpen Access PDF

Abstract

The premise that subjects must overcome the internal resistance of a DPI to achieve a peak inspiratory flow rate of 60 L min-1 12 for effective powder fluidization and deagglomeration of inhaled bronchodilators for the treatment of COPD is fundamentally incorrect. This reliance on a 60 L min-1 13 14 minimum flow leads to a significant overestimate of the percentage of patients with suboptimal flow 15 rates, and this criterion should not be used as a biomarker to guide transition of patients to an active 16 device. Alternatively, the minimum threshold energy for effective drug delivery with passive dry 17 powder inhalers (DPI) occurs at peak inspiratory pressures greater than or equal to about 1 kPa, 18 independent of the internal resistance of the device. Most patients can achieve the threshold 1 kPa 19 pressure drop even during periods of acute wheeze.

Topics & Concepts

Dry-powder inhalerMedicineWheezeVolumetric flow ratePressure dropFlow resistanceAirflowMaterials scienceFlow (mathematics)AnesthesiaInternal medicineMathematicsInhalerAsthmaMechanicsThermodynamicsPhysicsGeometryInhalation and Respiratory Drug DeliveryAsthma and respiratory diseasesIntravenous Infusion Technology and Safety