Development and validation of the RP-HPLC method for quantification of tavaborole
Shiv Kumar Prajapati, Ankit Jain, Meenakshi Bajpai
Abstract
) of 3.43 was observed, indicating significant interaction and good separation. Forced degradation (FD) or stress tests were performed for chemical and physical photolytic stress conditions, and the results observed were within the specified limits. The stability in the analytical solution was observed for up to 35 hours at 5 °C, confirming the stability of the solution. Validation of the developed HPLC method confirmed the system's suitability, precision, linearity, accuracy, FD, robustness, and results. All validation criteria for the technique were within acceptable limits.
Topics & Concepts
ChromatographyComputer scienceChemistryAnalytical Methods in PharmaceuticalsAnalytical Chemistry and Sensors