Litcius/Paper detail

Top-Down Characterization and Intact Mass Quantitation of a Monoclonal Antibody Drug from Serum by Use of a Quadrupole TOF MS System Equipped with Electron-Activated Dissociation

John F. Kellie, Nicole A. Schneck, Jason Causon, Takashi Baba, John T. Mehl, Kerstin I. Pohl

2022Journal of the American Society for Mass Spectrometry23 citationsDOI

Abstract

Time-of-flight MS systems for biopharmaceutical and protein characterization applications may play an even more pivotal role in the future as biotherapeutics increase in drug pipelines and as top-down MS approaches increase in use. Here, a recently developed TOF MS system is examined for monoclonal antibody (mAb) characterization from serum samples. After immunocapture, purified drug material spiked into monkey serum or dosed for an in-life study is analyzed by top-down MS. While characterization aspects are a distinct advantage of the MS platform, MS system and software capabilities are also shown regarding intact protein quantitation. Such applications are demonstrated to help enable comprehensive protein molecule quantitation and characterization by use of TOF MS instrumentation.

Topics & Concepts

ChemistryMonoclonal antibodyQuadrupole time of flightChromatographyCharacterization (materials science)BiopharmaceuticalMass spectrometryAntibodyTandem mass spectrometryNanotechnologyGeneticsBiologyImmunologyMaterials scienceMass Spectrometry Techniques and ApplicationsAnalytical chemistry methods developmentIon-surface interactions and analysis