Instability of rotational activation as atrial fibrillation drivers: Assessment by ExTRa Mapping system
Tetsuma Kawaji, Takanori Aizawa, Shun Hojo, Hidenori Yaku, Kenji Nakatsuma, Kazuhisa Kaneda, Masashi Kato, Takafumi Yokomatsu, Shinji Miki
Abstract
BACKGROUND: ExTRa Mapping™ has developed to visualize rotational activation as atrial fibrillation (AF) drivers. The current study was sought to evaluate the instability of AF drivers by ExTRa Mapping™. METHODS: Variation of nonpassively activated ratio (%NP) among three-time repetitive recordings before and after pulmonary vein isolation (PVI) in left atrium was assessed in 26 persistent AF patients. The recoding time was set at 5 or 8 s for the respective patients. The outcome measures included %NP at each recording, mean value of the three-time recordings, and the instability index, which was defined as maximum difference per mean %NP × 100 (%). RESULTS: Total 683 sites 2049 recordings were assessed. Mean %NP was 33.3(23.3-42.7)%, and higher in sites with severe (≥50%) and patchy low voltage area than those without, but not in those with severe complex fractionated atrial electrogram area. There was significant correlation between actual and mean %NP (R = 0.86, P < .001), but maximum difference among the repetitive recordings was 16(10-24)%. The instability index of %NP was 55.9(30.9-83.6)%, and significantly lower at the recordings of 8 s compared with 5 s (50.6[28.6-78.4]% vs. 60.4[35.0-90.0]%, P = .004). Furthermore, it was higher at sites with lower reliability of the recordings. After PVI, mean %NP significantly decreased (28.7[18.3-36.7]% vs. 37.7[28.7-45.7]%, P < .001), but the instability index significantly increased compared with those before PVI (60.0[35.0-92.7]% vs. 48.9[29.1-75.0]%, P = .001). CONCLUSION: Rotational activation as AF drivers assessed by ExTRa Mapping™ is unstable, and repetitive and longer recording is required for the reliable assessment even after PVI.