Evaluation of a Novel Non-Diffractive Extended Depth of Focus Intraocular Lens – First Results from a Prospective Study
Volkan Tahmaz, Sebastian Siebelmann, Konrad R. Koch, Claus Cursiefen, Achim Langenbucher, Robert Hoerster
Abstract
PURPOSE: To evaluate a novel hydrophobic, non-diffractive, extended depth of focus (EDOF) intraocular lens (IOL) design in comparison to two monofocal aspheric lenses. METHODS: Inclusion criteria for this prospective, monocentric cohort study were opacification of the crystalline lens and patients' wishes for surgery. In the case of the EDOF IOL, patients asked for a presbyopia correction. All patients received surgery on both eyes. Corrected and uncorrected distance visual acuity (CDVA, UCDVA), uncorrected and distance corrected intermediate visual acuity (UIVA, DCIVA) and defocus curves (all monocular and binocular) were compared three months postoperatively. RESULTS: < 0.0001) with the EDOF IOL. Defocus curves showed a depth of focus at 20/23 Snellen of 1.6 vs. 0.83 diopters (D) in the EDOF IOL, vs. the monofocal IOL. No patient reported halos or starbursts in non-standardized questioning. CONCLUSION: This non-diffractive EDOF IOL provided comparably high UCDVA and significantly higher DCIVA than the mono-focal lenses, causing only mild optical phenomena.