Litcius/Paper detail

A real‐world study of user characteristics, safety and efficacy of open‐source closed‐loop systems and Medtronic 670G

Roshell Jeyaventhan, Geraldine Gallen, Pratik Choudhary, Sufyan Hussain

2021Diabetes Obesity and Metabolism37 citationsDOIOpen Access PDF

Abstract

We report a real-world evaluation of the first commercially approved automated insulin delivery (AID) system, MiniMed 670G (670G), and open source-automated insulin delivery (OS-AID) systems. This was undertaken as a retrospective observational study in adults with type 1 diabetes using AID systems for 6 months or longer in a publicly funded health service using clinically validated data. Sixty-eight adults (38 670G, 30 OS-AID systems) were included. OS-AID system users were younger, had a shorter diabetes duration and a higher education status. OS-AID systems displayed a significantly better change in HbA1c (median -0.9% [-0.4%, -1.1%] vs. -0.1% [IQR -0.7%, 0.2%], P = .004) and time in range 3.9-10 mmol/L (mean 78.5%, SD ± 12.0% vs. 68.2% ± 14.7%, P = .024) compared with 670G. Both systems showed minimal hypoglycaemia, with OS-AID systems revealing significantly improved secondary outcomes of mean glucose and percentage of time more than 10 mmol/L, with a higher percentage of time of less than 3 mmol/L. OS-AID system users displayed improved glycaemic outcomes with no clinical safety concerns compared with 670G, although higher weight-adjusted insulin dose and weight gain were noted. The study highlights key differences in OS-AID system user characteristics that are important for interpreting real-world findings from recent OS-AID system studies.

Topics & Concepts

Insulin deliveryMedicineClosed loopObservational studyDelivery systemType 1 diabetesDiabetes mellitusInsulinInternal medicineEndocrinologyBiomedical engineeringEngineeringControl engineeringDiabetes Management and ResearchPancreatic function and diabetesDiabetes and associated disorders
A real‐world study of user characteristics, safety and efficacy of open‐source closed‐loop systems and Medtronic 670G | Litcius