Evaluation of three biometric devices: ocular parameters and calculated intraocular lens power

Rivkah Lender, Devora Mirsky, Riki Greenberger, Zipora Boim, Lee Ben-Yaakov, Chaya Kashtan, Ibrahim Naffar, Shira Shine, Itay Chowers, Hadas Ben‐Eli

2022Scientific Reports19 citationsDOIOpen Access PDF

Abstract

Cataract surgery is among the most common medical procedures, and accurate ocular biometry measurements are key for successful visual outcome. The current study evaluated data obtained by the Eyestar 900, Anterion, IOLMaster700 biometers and the Pentacam corneal topographer. Compared values were axial length (AL), anterior chamber depth (ACD), steep- and flat-K, cylinder and axis. Clinical impact was assessed by calculating intraocular lens (IOL) power using the mean values of every parameter and the Barrett and Kane formulas, stratified by device and amount of cylinder. IOL was re-calculated for each device substituting Pentacam K-values. This study included 196 eyes (98 participants) of cataract surgery candidates. When comparing the IOLMaster to the Eyestar (157 eyes), no difference was found in mean AL or ACD measurements (P > 0.05). Steep-K measurements differed between these devices and the Pentacam (P = 0.01). AL and ACD measurements differed between the IOLMaster and Anterion (38 eyes; P < 0.05). Strong correlations (range 0.72-0.99) were found between all four devices. Bland-Altman analysis demonstrated excellent agreement between biometry devices other than ACD between the IOLMaster and Eyestar. Calculated IOL power was 0.50-1.00 diopter (D) lower with the IOLMaster. Cylinder power was 0.75D higher in all biometers when Pentacam K-values were substituted.

Topics & Concepts

OphthalmologyDioptreIntraocular lensCataract surgeryLimits of agreementMedicineOptometryMathematicsOpticsPhysicsVisual acuityNuclear medicineCorneal surgery and disordersOphthalmology and Visual Impairment StudiesIntraocular Surgery and Lenses
Evaluation of three biometric devices: ocular parameters and calculated intraocular lens power | Litcius