Biomechanical Evaluation of Different Plate Configurations for Midshaft Clavicle Fracture Fixation

Joep Kitzen, Kent Paulson, Robert Korley, Paul Duffy, Claude R. Martin, Prism Schneider

2022JBJS Open Access16 citationsDOIOpen Access PDF

Abstract

Background: Dual-plate constructs have become an increasingly common fixation technique for midshaft clavicle fractures and typically involve the use of mini-fragment plates. The goal of this technique is to reduce plate prominence and implant irritation. However, limited biomechanical data exist for these lower-profile constructs. The study aim was to compare dual mini-fragment orthogonal plating with small-fragment clavicle plates for biomechanical noninferiority and to determine if an optimal plate configuration could be identified using a cadaveric model. Methods: Twenty-four cadaveric clavicles were randomized to 1 of 6 groups, stratified by computed tomography-based bone mineral content (BMC): precontoured superior or anterior fixation using a single 3.5-mm Locking Compression Plate (LCP), and 4 different dual-plating constructs utilizing 2.4-mm and 2.7-mm Adaptation plates or LCPs. An inferior butterfly fracture was created. Axial, torsional, and bending (anterior and superior surface loading) stiffnesses were determined through nondestructive cyclic testing, followed by a load-to-failure test in 3-point superior surface bending. Results: For axial stiffness, the 2 dual-plate constructs with a superior 2.4-mm and anterior 2.7-mm plate (either Adaptation or LCP) were significantly stiffer than the other 4 constructs (p = 0.021 and p = 0.034). For both superior and anterior bending, the superior 2.4-mm and anterior 2.7-mm plate constructs were significantly stiffer when compared with the 3.5-mm superior plate (p = 0.043). No significant differences were found in torsional stiffness or load to failure between the different constructs. Conclusions: Dual plating using mini-fragment plates is biomechanically superior for the fixation of midshaft clavicle fractures when compared with a single, superior, 3.5-mm plate and has biomechanical properties similar to those of a 3.5-mm plate placed anteriorly. With the exception of axial stiffness, no significant differences were found when different dual-plating constructs were compared with each other. Clinical Relevance: This study validates the use of dual plating for midshaft clavicle fractures.

Topics & Concepts

Cadaveric spasmFixation (population genetics)ClavicleMaterials scienceStiffnessBendingOrthodonticsBending stiffnessBiomedical engineeringAnatomyMedicineComposite materialPopulationEnvironmental healthShoulder and Clavicle InjuriesTrauma Management and DiagnosisShoulder Injury and Treatment