Litcius/Paper detail

Endovascular Baroreflex Amplification With the MobiusHD Device in Patients With Heart Failure and Reduced Ejection Fraction: Interim Analysis of the First-in-Human Results

Kerstin Piayda, Kolja Sievert, Horst Sievert, Tamaz Shaburishvili, Irakli Gogorishvili, Martin T. Rothman, James L. Januzzi, JoAnn Lindenfeld, Gregg W. Stone

2022Structural Heart12 citationsDOIOpen Access PDF

Abstract

Background: Endovascular baroreflex amplification with the MobiusHD, a self-expanding stent-like device that is implanted in the internal carotid artery, was designed to reduce the sympathetic overactivity that contributes to progressive heart failure with reduced ejection fraction. Methods: Symptomatic patients (New York Heart Association class III) with heart failure with reduced ejection fraction (left ventricular ejection fraction [LVEF] ≤40%) despite guideline directed medical therapy and n-terminal pro-B type natriuretic peptide (NT-proBNP) levels ≥400 pg/mL in whom carotid ultrasound and computed tomographic angiography demonstrated absence of carotid plaque were enrolled. Baseline and follow-up measures included 6-minute walk distance (6MWD), Kansas City Cardiomyopathy Questionnaire overall summary score (KCCQ OSS), and repeat biomarkers and transthoracic echocardiography. Results: Twenty-nine patients underwent device implantation. The mean age was 60.6 ± 11.4 years, and all had New York Heart Association class III symptoms. Mean KCCQ OSS was 41.4 ± 12.7, mean 6MWD was 216.0 ± 43.7 m, median NT-proBNP was 1005.9 pg/mL (894, 1294), and mean LVEF was 34.7 ± 2.9%. All device implantations were successful. Two patients died (161 days and 195 days) and one stroke occurred (170 days) during follow-up. For the 17 patients with 12-month follow-up, mean KCCQ OSS improved by 17.4 ± 9.1 points, mean 6MWD increased by 97.6 ± 51.1 meters, a mean 28.4% reduction from the baseline NT-proBNP concentration was found, and mean LVEF improved by 5.6% ± 2.9 (paired data). Conclusion: Endovascular baroreflex amplification with the MobiusHD device was safe and effected positive changes in quality of life, exercise capacity, and LVEF, consistent with observed reductions in NT-proBNP levels.

Topics & Concepts

Ejection fractionMedicineHeart failureCardiologyInternal medicineStroke volumeBaroreflexBrain natriuretic peptideHeart rateBlood pressureHeart Rate Variability and Autonomic ControlVagus Nerve Stimulation ResearchAtrial Fibrillation Management and Outcomes
Endovascular Baroreflex Amplification With the MobiusHD Device in Patients With Heart Failure and Reduced Ejection Fraction: Interim Analysis of the First-in-Human Results | Litcius