Reduced Intravenous Contrast Dose Portal Venous Phase Photon-Counting Computed Tomography Compared With Conventional Energy-Integrating Detector Portal Venous Phase Computed Tomography

Bari Dane, Tarub S. Mabud, Kira Melamud, Luke Ginocchio, Paul Smereka, Mabel Okyere, Thomas O’Donnell, Alec J. Megibow

2024Journal of Computer Assisted Tomography10 citationsDOI

Abstract

OBJECTIVE: The aim of this study was to compare portal venous phase photon-counting CT (PCCT) using 20 cc less than weight-based contrast dosing with energy-integrating detector CT (EID-CT) using weight-based dosing by quantitative and qualitative analysis. METHODS: Fifty adult patients who underwent a reduced intravenous contrast dose portal venous phase PCCT from May 1, 2023, to August 10, 2023, and a prior portal-venous EID-CT with weight-based contrast dosing were retrospectively identified. Hounsfield units (HU) and noise (SD of HU) were obtained from region-of-interest measurements on 70-keV PCCT and EID-CT in 4 hepatic segments, the main and right portal vein, and both paraspinal muscles. Signal-to-noise and contrast-to-noise ratios were computed. Three abdominal radiologists qualitatively assessed overall image quality, hepatic enhancement, and confidence for metastasis identification on 5-point Likert scales. Readers also recorded the presence/absence of hepatic metastases. Quantitative variables were compared with paired t tests, and multiple comparisons were accounted for with a Bonferroni-adjusted α level of .0016. Ordinal logistic regression was used to evaluate qualitative assessments. Interreader agreement for hepatic metastases was calculated using Fleiss' κ. RESULTS: Fifty patients (32 women; mean [SD] age, 64 [13] years) were included. There was no significant difference in hepatic HU, portal vein HU, noise, and signal-to-noise or contrast-to-noise ratio between reduced contrast dose portal venous phase PCCT versus EID-CT (all P s > 0.0016). Image quality, degree of hepatic enhancement, and confidence for metastasis identification were not different for reduced dose PCCT 70-keV images and EID-CT ( P = 0.06-0.69). κ Value for metastasis identification was 0.86 (95% confidence interval, 0.70-1.00) with PCCT and 0.78 (95% confidence interval, 0.59-0.98) with EID-CT. CONCLUSION: Reduced intravenous contrast portal venous phase PCCT 70-keV images had similar attenuation and image quality as EID-CT with weight-based dosing. Metastases were identified with near-perfect agreement in reduced dose PCCT 70-keV images.

Topics & Concepts

MedicineNuclear medicineRadiologyHounsfield scaleConfidence intervalComputed tomographyInternal medicineAdvanced X-ray and CT ImagingRadiation Dose and ImagingAdvanced X-ray Imaging Techniques