Preliminary Experience of 5.0 T Higher Field Abdominal Diffusion‐Weighted <scp>MRI</scp>: Agreement of Apparent Diffusion Coefficient With 3.0 T Imaging
Yunfei Zhang, Chun Yang, Liang Liang, Zhang Shi, Shuo Zhu, Caizhong Chen, Yongming Dai, Mengsu Zeng
Abstract
BACKGROUND: Recently, a prototype 5.0 T whole-body MRI scanner was developed. A 5.0 T diffusion-weighted imaging (DWI) may help overcome the issues that limit 3.0 T DWI. PURPOSE: To evaluate the feasibility of 5.0 T high-field DWI in the upper abdomen and assess the agreement of the apparent diffusion coefficient (ADC) with that from 3.0 T abdominal DWI. STUDY TYPE: Prospective proof of concept. POPULATION: Nine volunteers (mean ± SD age: 37.3 ± 7.0 years, 8 M), eight healthy and one with liver and kidney cysts. FIELD STRENGTH/SEQUENCE: 3.0 T and 5.0 T; respiratory-triggered spin-echo echo-planar-imaging (SE-EPI)-based DWI sequence. ASSESSMENT: Subjective image quality scores. The ADC values in abdominal organs (liver, pancreas, spleen, and kidney) were measured by two observers for evaluating the interobserver and interfield agreement. STATISTICAL TESTS: Wilcoxon-rank sum test, Bland-Altman analysis, intraclass correlation coefficients (ICCs), and coefficients of variation (CVs). RESULTS: The 5.0 T DWI displayed an increase in subjective image quality score compared to 3.0 T DWI without the significant difference (3.0 T DWI: 3.50 ± 0.47, 5.0 T DWI: 3.72 ± 0.42, P = 0.157). Both the interfield and interobserver agreements of ADC values were substantial to excellent (ICCs = 0.640-0.902). For all four upper abdominal organs, there were no significant differences between the ADC values measured by two observers and between the ADC values of 3.0 T and 5.0 T DWI (P = 0.134-1.000). The CVs of ADC measurements from 3.0 T and 5.0 T DWI were all less than 15.0% (6.7%-14.2%). DATA CONCLUSION: The substantial to excellent agreements between the ADC values measured with 3.0 T and 5.0 T DWI for liver, pancreas, spleen, and kidney suggested that 5.0 T DWI can be applied for abdominal imaging. The ADC values from 5.0 T abdominal DWI hold the potential to serve as the quantitative markers for clinical investigations. EVIDENCE LEVEL: 2 TECHNICAL EFFICACY: Stage 1.