Prospective study of 1308 nasopharyngeal swabs from 1033 patients using the LUMIPULSE SARS-CoV-2 antigen test: Comparison with RT-qPCR

Yosuke Hirotsu, Makoto Maejima, Masahiro Shibusawa, Kenji Amemiya, Yuki Nagakubo, Kazuhiro Hosaka, Hitomi Sueki, Miyoko Hayakawa, Hitoshi Mochizuki, Toshiharu Tsutsui, Yumiko Kakizaki, Yoshihiro Miyashita, Masao Omata

2021International Journal of Infectious Diseases60 citationsDOIOpen Access PDF

Abstract

BACKGROUND: Reverse transcription polymerase chain reaction (RT-PCR) is the gold standard for detection of severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2). Previously, the accuracy of the quantitative LUMIPULSE SARS-CoV-2 antigen test was demonstrated using samples collected retrospectively. In this study, the LUMIPULSE antigen test was clinically validated using prospective samples. METHODS: In total, 1033 nasopharyngeal swab samples were collected from 1033 individuals, and an additional 275 follow-up samples were collected from 43 patients who subsequently tested positive for coronavirus disease 2019 (COVID-19). All 1308 samples were subjected to quantitative RT-PCR (RT-qPCR) and the antigen test. The antibody response was investigated for patients with discordant results to clarify if seroconversion had occurred. RESULTS: RT-qPCR identified 990 samples as negative and 43 as positive, while the antigen test identified 992 samples as negative, 37 as positive and four as inconclusive. The overall concordance rate was 99.7% (1026/1029). Sensitivity, specificity, positive predictive value and negative predictive value of the antigen test were 92.5% (37/40), 100% (989/989), 100% (37/37) and 99.7% (989/992), respectively, after exclusion of the four inconclusive results. The kappa coefficient was 0.960 (95% confidence interval 0.892-0.960), suggesting excellent agreement between the two tests. Seropositivity in five of seven patients with discordant results suggested that the discrepancy was caused by samples collected during the late phase of infection. Using follow-up samples, correlation was observed between the antigen level and the viral load or cycle threshold value. The concordance rate between these test results tended to be high among samples collected 0-9 days after symptom onset, but this decreased gradually in samples collected thereafter. CONCLUSIONS: This prospective study demonstrated that the LUMIPULSE antigen test is a highly accurate diagnostic test for SARS-CoV-2.

Topics & Concepts

ConcordanceMedicineAntigenSeroconversionInternal medicineConfidence intervalImmunologyReal-time polymerase chain reactionGold standard (test)Prospective cohort studyPredictive value of testsGastroenterologyAntibodyBiologyGeneBiochemistrySARS-CoV-2 detection and testingSARS-CoV-2 and COVID-19 ResearchBiosensors and Analytical Detection
Prospective study of 1308 nasopharyngeal swabs from 1033 patients using the LUMIPULSE SARS-CoV-2 antigen test: Comparison with RT-qPCR | Litcius