Litcius/Paper detail

A novel energy sequence optimization algorithm for efficient spot-scanning proton arc (SPArc) treatment delivery

Gang Liu, Xiaoqiang Li, Lewei Zhao, Weili Zheng, An Qin, Sheng Zhang, Craig Stevens, Di Yan, Peyman Kabolizadeh, Xuanfeng Ding

2020Acta Oncologica51 citationsDOIOpen Access PDF

Abstract

Background Spot-scanning proton arc therapy (SPArc) has been proposed to improve dosimetric outcome and to simplify treatment workflow. To efficiently deliver a SPArc plan, it’s crucial to minimize the number of energy layer switches (ELS) a sending because of the magnetic hysteresis effect. In this study, we introduced a new SPArc energy sequence optimization algorithm (SPArc_seq) to reduce ascended ELS and to investigate its impact on the beam delivery time (BDT).Method and materials An iterative energy layer sorting and re-distribution mechanism following the direction of the gantry rotation was implemented in the original SPArc algorithm (SPArc_orig). Five disease sites, including prostate, lung, brain, head neck cancer (HNC) and breast cancer were selected to evaluate this new algorithm. Dose-volume histogram (DVH) and plan robustness were used to assess the plan quality for both SPArc_seq and SPArc_orig plans. The BDT evaluations were analyzed through two methods: 1. fixed gantry angle delivery (BDTfixed) and 2. An in-house dynamic arc scanning controller simulation which considered of gantry rotation speed, acceleration and deceleration (BDTarc).Results With a similar total number of energy layers, SPArc_seq plans provided a similar nominal plan quality and plan robustness compared to SPArc_orig plans. SPArc_seq significantly reduced the number of ascended ELS by 83% (19 vs.115), 70% (16 vs. 64), 82% (19 vs. 104), 80% (19 vs. 94) and 70% (9 vs. 30), which effectively shortened the BDTfixed by 65% (386 vs. 1091 s), 61% (235 vs. 609 s), 64% (336 vs. 928 s), 48% (787 vs.1521 s) and 25% (384 vs. 511 s) and shortened BDTarc by 54% (522 vs.1128 s), 52% (310 vs.645 s), 53% (443 vs. 951 s), 49% (803 vs.1583 s) and 26% (398 vs. 534 s) in prostate, lung, brain, HNC and breast cancer, respectively.Conclusions The SPArc_seq optimization algorithm could effectively reduce the BDT compared to the original SPArc algorithm. The improved efficiency of the SPArc_seq algorithm has the potential to increase patient throughput, thereby reducing the operation cost of proton therapy.

Topics & Concepts

AlgorithmRobustness (evolution)Nuclear medicinePencil-beam scanningProton therapyMedicineComputer scienceRadiation therapySurgeryBiochemistryGeneChemistryRadiation Therapy and DosimetryAdvanced Radiotherapy TechniquesBoron Compounds in Chemistry
A novel energy sequence optimization algorithm for efficient spot-scanning proton arc (SPArc) treatment delivery | Litcius