Active modulation of intracavity laser intensity with the Pound–Drever–Hall locking for photoacoustic spectroscopy

Zhen Wang, Haoyun Wei, Yan Li, Ruifeng Kan, Wei Ren

2020Optics Letters19 citationsDOI

Abstract

Here we report a novel, to the best of our knowledge, method of active intracavity intensity modulation for cavity-enhanced photoacoustic spectroscopy (PAS) without the need for any external optical modulators. Based on the Pound–Drever–Hall (PDH) locking technique, a dither is added to the PDH error signal to periodically vary the locking point between the laser frequency and optical cavity within a sub-MHz frequency range. While significantly enhancing the intracavity laser intensity, the optical cavity also acts as an intensity modulator. As a proof-of-principle, we demonstrated the PAS of <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" display="inline"> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:msub> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mi mathvariant="normal">C</mml:mi> </mml:mrow> <mml:mn>2</mml:mn> </mml:msub> </mml:mrow> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:msub> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mi mathvariant="normal">H</mml:mi> </mml:mrow> <mml:mn>2</mml:mn> </mml:msub> </mml:mrow> </mml:math> by placing a photoacoustic cell ( <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" display="inline"> <mml:mi>Q</mml:mi> </mml:math> -factor <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" display="inline"> <mml:mo>∼</mml:mo> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mn>10</mml:mn> </mml:mrow> </mml:math> ) inside a Fabry–Perot cavity (finesse <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" display="inline"> <mml:mo>∼</mml:mo> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mn>628</mml:mn> </mml:mrow> </mml:math> ) and adopting the proposed intracavity intensity modulation scheme. By detecting the weak <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" display="inline"> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:msub> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mi mathvariant="normal">C</mml:mi> </mml:mrow> <mml:mn>2</mml:mn> </mml:msub> </mml:mrow> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:msub> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mi mathvariant="normal">H</mml:mi> </mml:mrow> <mml:mn>2</mml:mn> </mml:msub> </mml:mrow> </mml:math> line at <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" display="inline"> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mn>6412.73</mml:mn> </mml:mrow> <mml:mspace width="thickmathspace"/> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:msup> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mi mathvariant="normal">c</mml:mi> <mml:mi mathvariant="normal">m</mml:mi> </mml:mrow> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mo>−</mml:mo> <mml:mn>1</mml:mn> </mml:mrow> </mml:msup> </mml:mrow> </mml:math> , the sensor achieves a normalized noise equivalent absorption (NNEA) coefficient of <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" display="inline"> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mn>1.5</mml:mn> </mml:mrow> <mml:mo>×</mml:mo> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:msup> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mn>10</mml:mn> </mml:mrow> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mo>−</mml:mo> <mml:mn>11</mml:mn> </mml:mrow> </mml:msup> </mml:mrow> <mml:mspace width="thickmathspace"/> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:msup> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mi mathvariant="normal">c</mml:mi> <mml:mi mathvariant="normal">m</mml:mi> </mml:mrow> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mo>−</mml:mo> <mml:mn>1</mml:mn> </mml:mrow> </mml:msup> </mml:mrow> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:msup> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mi mathvariant="normal">W</mml:mi> <mml:mi mathvariant="normal">H</mml:mi> <mml:mi mathvariant="normal">z</mml:mi> </mml:mrow> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mo>−</mml:mo> <mml:mn>1</mml:mn> <mml:mrow class="MJX-TeXAtom-ORD"> <mml:mo>/</mml:mo> </mml:mrow> <mml:mn>2</mml:mn> </mml:mrow> </mml:msup> </mml:mrow> </mml:math> . This method enables the continuous locking of laser frequency and optical cavity, and it achieves the intracavity intensity modulation with an adjustable modulation depth as well.

Topics & Concepts

OpticsPhotoacoustic spectroscopyMaterials scienceLaserSpectroscopyModulation (music)OptoelectronicsPhysicsQuantum mechanicsAcousticsSpectroscopy and Laser ApplicationsLaser Design and ApplicationsAtomic and Subatomic Physics Research
Active modulation of intracavity laser intensity with the Pound–Drever–Hall locking for photoacoustic spectroscopy | Litcius