Validation of PRIMO monte carlo model of clinac<sup>®</sup>ix 6mv photon beam

B Sarin, B. Bindhu, B Saju, Raguram K Nair

2020Journal of Medical Physics12 citationsDOIOpen Access PDF

Abstract

PURPOSE: iX linear accelerator using PRIMO Monte Carlo (MC) code and to assess PRIMO as an independent MC-based dose verification and quality assurance tool. MATERIALS AND METHODS: Acuros XB dose calculations in the case of both homogenous and inhomogeneous mediums. The gamma analysis method with the acceptance criteria of 2%, 2 mm was used for the comparison of dose distributions. RESULTS: XB shows a minimum gamma analysis pass rate of 99%. CONCLUSIONS: This study shows that the research software PRIMO can be used as a treatment planning system-independent quality assurance and dose verification tool in daily clinical practice. Further validation will be performed with different energies, complex multileaf collimators fields, and with dynamic treatment fields.

Topics & Concepts

Quality assuranceMonte Carlo methodLinear particle acceleratorBeam (structure)Nuclear medicinePhysicsPhotonRadiation treatment planningComputational physicsMedical physicsNuclear engineeringMathematicsOpticsRadiation therapyStatisticsEngineeringMedicineExternal quality assessmentInternal medicineOperations managementAdvanced Radiotherapy TechniquesAdvanced X-ray and CT ImagingRadiation Dose and Imaging