Litcius/Paper detail

Impact of Multileaf Collimator Width on Dose Distribution in HyperArc Fractionated Stereotactic Irradiation for Multiple (-) Brain Metastases

Shingo Ohira, Yoshihiro Ueda, Naoyuki Kanayama, Masaru Isono, Shoki Inui, Riho Komiyama, Hayate Washio, Masayoshi Miyazaki, Masahiko Koizumi, Teruki Teshima, Koji Konishi

2021Anticancer Research24 citationsDOIOpen Access PDF

Abstract

Background/Aim: To assess the impact of the width of multileaf collimator (MLC) on dose distributions on HyperArc fractionated stereotactic irradiation for multiple (5-10) brain metastases. Patients and Methods: Twenty-one HyperArc (HA) plans were generated using the high definition (HD) MLC (2.5 mm) to deliver 30-35 Gy in 3-5 fractions (HA-HD). The HyperArc plans using Millennium (ML) MLC (5 mm) were retrospectively generated (HA-ML) using the same planning parameters with HA-HD. Dosimetric parameters between the planning target volume (PTV) and organs at risk (OARs) were compared. Results: The conformity index was significantly higher (p&lt;0.0001) in the HA-HD plans (0.95±0.04) than that in the HA-ML plans (0.92±0.06). The HA-HD provided significantly lower (p&lt;0.0001) gradient index (5.6±2.5) than HA-ML (6.2±3.5). For the brainstem and retina (right), a statistically significant difference (p&lt;0.05) was observed between the HA-HD (12.8±10.9 and 2.8±1.7 Gy, for brainstem and retina, respectively) and HA-ML (13.6±11.1 and 3.0±1.8 Gy) plans. For the brain tissue, the HA-HD plans statistically significantly reduced dosimetric parameters (p&lt;0.0001) in all evaluated dose range (V<sub>6Gy</sub>-V<sub>28Gy</sub>). Conclusion: The narrower MLC provided significantly higher conformity, steeper dose gradient, and better normal tissue sparing.

Topics & Concepts

Nuclear medicineMultileaf collimatorMedicineStereotactic radiotherapyRadiosurgeryRadiation therapyDosimetryRadiation treatment planningRadiologyAdvanced Radiotherapy TechniquesRadiation Therapy and DosimetryBrain Metastases and Treatment
Impact of Multileaf Collimator Width on Dose Distribution in HyperArc Fractionated Stereotactic Irradiation for Multiple (-) Brain Metastases | Litcius