Clinical and radiological outcomes of a stemless reverse shoulder implant: a two-year follow-up in 56 patients
Christian Schoch, Johannes E. Plath, Leander Ambros, Michaël Geyer, Michael Dittrich
Abstract
BACKGROUND: Since the introduction of stemless anatomic shoulder arthroplasty, many studies have been published on anatomic implants. For reverse stemless implants, however, there are only a few clinical follow-up studies available. The current clinical case series aims to present clinical and radiological outcomes of a new stemless reverse prosthesis system (Lima Shoulder Modular Replacement stemless). METHODS: We prospectively evaluated the outcome of 56 stemless total shoulder arthroplasties in 56 patients with a mean age of 61.2 years (46-76 years) at the time of implantation at a minimum follow-up of 24 months (range 24-41 months). All patients were physically and radiologically examined. Clinical outcomes were evaluated by using the Constant-Murley Score and the Subjective Shoulder Value. RESULTS: < .001) and active range of motion (abduction 72° to 130°, flexion 36° to 138°, and external rotation 16°to 28°). There was one complete loosening of the humeral component without reoperation. Radiolucency lines were observed in anteroposterior or axial radiographs at the humeral component in 23% of the cases, most of them in anteroposterior view at the calcar region. Radiolucency line findings did not affect clinical outcomes. Major complications or revisions did not occur so far. CONCLUSION: At short-term follow-up, stemless reverse shoulder systems show comparable clinical and radiological outcomes compared to stemmed reverse implants in the literature.