Feasibility study of a novel wireless localization technique using radiofrequency identification markers for small and deeply located lung lesions
Yojiro Yutaka, Toshihiko Satō, Satona Tanaka, So Miyahara, Akihiro Yoshizawa, Satoshi Morita, Hiroshi Date
Abstract
Objectives: To evaluate the safety and efficacy of a novel wireless localization technique that uses radiofrequency identification markers for small and deep lung lesions. Methods: Preliminary use of the device was retrospectively evaluated in 2 Japanese centers. Under general anesthesia, a marker was placed as close as possible to the tumor via computed tomography-guided bronchoscopy in a hybrid operation theater. Surgeons located the marker without lung palpation using a detection probe the tone of which changed to indicate the marker-probe distance. Efficacy was defined as functional marker placement (bronchoscopy time and marker position) and deep margin distance. Results: Twelve markers were placed for 11 lesions (mean size, 6.8 ± 2.7 mm) located at a mean depth from the pleura of 11.4 ± 8.4 mm (range = 0-26.0 mm). Of 12 markers, 7 markers (58.3%) were placed within 10 mm from the lesion in 25.5 ± 14.4 minutes. For the 11 wedge resections, markers were placed at a mean distance of 6.7 mm (range, 0-13.0 mm) from the lesion and a mean distance of 14.4 mm (range, 3.0-42.0 mm) from the pleura. All markers were recovered without complications, and all tumors were resected with negative margins. For 5 lesions >10 mm deep to the pleura (mean depth, 18.9 ± 5.5 mm; range, 11.0-26.0 mm), the median depth of the surgical margin was 11.6 ± 2.1 mm (range, 9.0-14.0 mm). Conclusions: Radiofrequency identification marking was safe and precisely localized small lung lesions, including their depth.