Litcius/Paper detail

An Ultrasensitive Lateral Flow Immunoassay Based on Metal-Organic Framework-Decorated Polydopamine for Multiple Sulfonylureas Adulteration in Functional Foods

Zixian He, Zhiwei Liu, Haihuan Xie, Pengjie Luo, Xiangmei Li

2023Foods15 citationsDOIOpen Access PDF

Abstract

Herein, an ultrasensitive lateral flow immunoassay (LFIA), based on metal-organic framework-decorated polydopamine (PCN-224@PDA) was first established to detect multiple sulfonylureas (SUs) in functional foods. The PCN-224@PDA was synthesized using the one-pot hydrothermal method and covalently coupled with SUs antibodies, and the coupling rate was up to 91.8%. The detection limits of the developed PCN-224@PDA-LFIA for multiple SUs in functional teas and capsules were 0.22-3.72 μg/kg and 0.40-3.71 μg/kg, and quantification limits were 0.75-8.19 μg/kg and 1.03-9.08 μg/kg, respectively. The analytical sensitivity was 128-fold higher than that of similar methods reported so far. The recovery rates ranged from 83.8 to 119.0%, with coefficients of variation of 7.6-14.4%. The parallel analysis of 20 real samples by LC-MS/MS confirmed the reliability of the proposed method. Therefore, our work offers novel, ultrasensitive, and rapid technical support for on-site monitoring of SUs in functional foods.

Topics & Concepts

Detection limitImmunoassayChromatographyChemistryMaterials scienceAntibodyBiologyImmunologyBiosensors and Analytical DetectionAdvanced biosensing and bioanalysis techniquesAnalytical chemistry methods development